7
•
Maintenance and Service
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 153
Emptying the Silicone Tubing
NOTICE
Always empty the silicone tubing when purging is complete. When the
purge valve is closed, purge liquid remains in the tubing connected to
the purge outlet. Leaving the liquid in the tubing for a longer period
limits the tubing lifetime.
1. To absorb any liquid that might be leaving the purge outlet when the
tubing is removed, place the tissue under the outlet.
2. On the purge/sensor unit, remove the silicone tubing from the purge
outlet. The other end of the tubing remains in the purge drain port
outlet.
3. Hold the tubing so that any liquid that is present in the tubing flows
off through the drain port outlet.
4. On the purge/sensor unit, consider reconnecting the silicone tubing.
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...