5
•
Installation
Page 76
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Figure 24: Seal wash connections
No.
Description
A
Single split sampler:
Seal wash connections on the top enclosure frame
B
Dual split sampler:
Seal wash connections on the right enclosure frame
1
Seal wash inlet port
1a
Seal wash line to reservoir
2
Seal wash outlet
2a
Seal wash line to pump
4. In the pump, connect the seal wash line to the tubing connector on
the free end of the peristaltic tubing.
Routing the Seal Wash Line to the Solvent Rack
1. Locate the seal wash line for connection from the autosampler to
the seal wash reservoir.
2. Connect the seal wash line on the autosampler:
¨
Single split sampler:
Connect the line to the seal wash inlet port
on the top enclosure frame (right tubing connector).
¨
Dual split sampler:
Connect the line to the seal wash inlet port
on the right enclosure frame (top tubing connector).
3. Route the seal wash line from the autosampler to the solvent rack
through the tubing guides in the autosampler and all modules above
the autosampler in the system stack.
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...