Troubleshooting
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 207
Message and Code
Description and Remedial Action
Code 4241 An autosampler is required.
To perform the procedure, an autosampler is required that is
unambiguously defined.
In the Chromeleon software, select the correct configuration for
your system and connect all devices.
Code 4242 The autosampler is offline.
The autosampler is not accessible via a System Interlink
connection.
Check that the autosampler is properly connected via System
Interlink and that the autosampler is powered on.
Code 4243 Unsupported autosampler.
The autosampler in the system configuration does not support
this procedure.
Code 4244 The column compartment is
offline.
The column compartment in the system configuration is not
accessible via the System Interlink connection.
Check that the column compartment is properly connected via
System Interlink and that the column compartment is powered
on.
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...