10
•
Modifying the Pump for Specific Applications
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 247
No.
Description
1
Mounting holes for mounting bracket (covered by push rivets while no
mounting bracket is installed in this position)
2
Mounting bracket
3
Mounting screws (Torx)
Follow these steps
1. Connect the capillary mixer on the purge/sensor unit. Mind the
direction of flow through the capillary mixer (indicated by the arrow
on the mixer).
TIP
To facilitate installation of the inline filter, loosely tighten the
capillary mixer on the purge/sensor unit so that you can easily access
the free end of the capillary.
2. Connect the inline filter to the capillary mixer. Mind the direction of
flow through the inline filter (indicated by the arrow on the filter).
3. Turn and insert the assembly of inline filter and capillary mixer into
the mounting bracket.
Figure 81: Inline filter with capillary mixer installed
No.
Description
1
Inline filter
2
Capillary mixer
4. Tighten the capillary mixer on the purge/sensor unit.
5. Reinstall the seal wash detector assembly in the pump (see
Replacing the Seal Wash Detector (
6. Reconnect the system pressure sensor cable.
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...