7
•
Maintenance and Service
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 185
3. Monitor the Audit Trail of the Instrument Configuration Manager
program to see whether the firmware update was successful or
failed.
¨
If the firmware update failed, turn the device off and on again
and repeat the firmware update.
¨
If the firmware update fails repeatedly, contact Thermo Fisher
Scientific Technical Support for assistance.
4. After a successful firmware update, requalification of the device may
be required. See the release notes for a recommendation.