Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
LED
Operational Status
Green, flashing – during
normal operation
In addition, the message "
Compression limit
reached
" (code 4156) appears in the Chromeleon Audit
Trail.
The compression value was 100% during each of the
past three strokes. If the compression value is lower
than 100% for several strokes afterward, the message
"
Compression back to normal
"(code 4157) will appear
in the Audit Trail.
For more information about the compression values,
see
Checking the Compression Values (
.
Green, flashing – during
pump maintenance
The pistons are in the appropriate position for piston
(seal) replacement.
Red, flashing
The pistons are not in the appropriate position for
normal operation (for example, after piston (seal)
replacement) or an error occurred during internal
maintenance (see
).
In both cases, an accompanying message appears in the
Chromeleon Audit Trail when you try to start the flow,
for example, "
The pump drive is still in undock
position; Please execute a dock command
" (code
4127). Follow the remedial actions for the
accompanying message in
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...