Safety
Page 26
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
2.4
Solvent and Additive Information
2.4.1
General Compatibility
To protect optimal functionality of the Vanquish system, observe these
recommendations on the use of solvents and additives:
•
The system must be used with reversed-phase (RP) compatible
solvents and additives only.
•
Use only solvents and additives that are compatible with all parts in
the flow path.
TIP
In a Vanquish Core system, normal-phase (NP) compatible solvents
and additives may be used if the VC-pumps and the VC-autosamplers are
modified with the components from the Normal-Phase (NP) kit. See
Normal-Phase Compatible Solvents and Additives (
NOTICE
If the system flow path includes a Vanquish charged aerosol detector,
observe the specific solvent and additive information for this detector.
Refer to the
Operating Manual
for the Vanquish charged aerosol
detector.
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...