7
•
Maintenance and Service
Page 130
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
7.5.1
Testing the Seal Wash System for Leakage
When
•
At regular intervals
•
If you performed maintenance on components in the pump head,
for example, after replacing the piston seals
Items required
•
Syringe (12 mL)
•
Seal wash liquid
Follow these steps
1. Disconnect the tubing from the seal wash detector.
Figure 32: Seal wash detector
No.
Description
1
Seal wash detector
2
Tubing connection port (port labeled
in
)
2. Insert a syringe into the open end of the tubing.
3. To allow the liquid to travel easily through the seal wash system,
press down the lever of the seal wash pump and hold it in that
position with one hand.
4. Pull out the syringe plunger to draw seal wash liquid into the tubing.
5. When about 10 mL of liquid are present in the syringe, release the
lever of the seal wash pump.
6. Press the lever firmly onto the tubing and press the liquid from the
syringe into the seal wash system.
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...