9
•
Pump-Specific Information
Page 224
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
9.2.5
Synchronizing the Injection Time with the Pump Strokes
To enhance the retention time precision with gradient applications, you
can synchronize the injection time of the Vanquish autosampler with the
strokes of the Vanquish pump. The synchronization ensures that all
injections are performed at the same phase of the pump cycle.
Synchronization is possible only when configured in the Chromeleon
software (refer to the
Chromeleon Help
).
Then, the following properties are available in the Chromeleon software:
•
SyncWithPump
To enable synchronization, set the property to
On
. To disable
synchronization, for example, for a specific application, set
SyncWithPump
to
Off
.
•
Pump
Shows the name of the pump to which the autosampler is linked.
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...