8
•
Troubleshooting
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 211
Follow these steps
1. In the Chromeleon software, check the
Compression
value for the
pump heads (=
CompressionDrv1
).
The values indicate the compression of the last stroke in percent.
The compression value should be less than 100%.
¨
With isocratic conditions, the compression value should remain
stable.
¨
When running a gradient, the compression value changes
roughly proportional to the pressure curve.
If the value is close to 100%, the required precompression may not
be reached and pulsation may occur.
The table shows guide values for some solvents (pure solvent,
degassed, delivered at 40 MPa) with the pump being purged and the
pressure being stable for some time. The values are linear to the
pressure.
Solvent
Compression in %
Water
25 % ± 10 %
Methanol
50 % ± 10 %
Acetonitrile
45 % ± 10 %
2. Take remedial action if required.
Compression
Remedial Action
Lower than indicated in the table
The outlet check valve may be defective.
Replace the valve cartridge. See
.
Higher than indicated in the table
•
The piston seal may be defective. Replace
the piston seal. See
•
The inlet check valve may be defective.
Replace the valve cartridge. See
.
Very high (message "
Compression
limit reached
" and low pressure)
•
Air bubbles may be present in the pump.
Make sure that no air bubbles are present in
the solvent lines and then, purge the pump.
•
If solvent shut off valves are installed:
Verify
that the valves are open.
•
Excessive leakage may be present on the
inlet check valve. Remove the solvent line
from the solvent reservoir and the solvent
line filter from the solvent line. While the
pump is running, observe the liquid in the
solvent line. If the liquid moves also against
the direction of flow during the stroke, the
inlet check valve may be defective. Replace
the valve cartridge. See
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...