LIT/5213/02   02/17

VIO

® 

200 S

2

The message box will say, “Please confirm the

settings by pressing any key!” (Note: the last

program used will be the one displayed). To

choose a different program: Prog. Setup 

 Select

program 

 scroll to Snare/Hot Biopsy.

QUICKSTART GUIDE for VIO

®

 200 S

Set up for Snare/Hot Biopsy Use

1

Turn ON the Unit.

3

Press the FOCUS Active Display button beside the

Monopolar receptacle.

4

Check the Program Screen to confirm the settings

(Default program shown).

Plug in the Active Cord and attach the Snare or 

Hot Biopsy Forceps.

5

6

Plug the Grounding Pad into the NEUTRAL

receptacle and place the Pad on the patient.

7

Place Footswitch at the physician’s feet. Verbally

confirm settings with physician.

8

READY to begin procedure.

Adjust settings per physician direction, increasing the 

Effect then max. Watts for more power.