7
•
Maintenance and Service
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 175
7.9.3
Replacing the Solvent Line Filter
When
•
Damage of solvent line filter
•
Impaired permeability of filter frit
Parts and additional items required
Figure 58: Solvent line filter
No.
Description
1
Solvent line filter, filter holder bottom part
2
Filter frit
3
Solvent line filter, filter holder top part
•
As applicable:
¨
Filter holder
¨
Filter frit
•
Solvent suitable for purging
Preparations
1. Turn off the pump flow and wait until the system pressure is down
to zero.
Follow these steps
1. Unscrew the cap of the solvent reservoir and remove the cap
together with the solvent line from the reservoir.
2. Remove the filter holder from the solvent line.
3.
As applicable
: Open the filter holder and remove the filter frit.
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...