Modifying the Pump for Specific Applications
Page 252
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Routine and Preventive Maintenance
•
Observe the information in
Routine and Preventive Maintenance
.
•
Depending on the application, slightly increased abrasion may be
observed with normal-phase piston seals. Therefore, consider
testing the static mixer for permeability at shorter intervals (see
Testing the Static Mixer for Permeability (
).
Testing the Pump for Leakage
Follow the instructions in
Testing the Pump for Leakage (General Test)
. With one exception, all values in the procedure that are
specific for the VC‑pump equally apply to the modified VC‑pump.
Exception
When the pressure is 55 MPa, apply the flow indicated in the table and
proceed with observing the pressure. The pressure should increase or
remain constant for a minimum of 1 minute. If you observe a drop in
pressure, find and eliminate the cause for the leak as appropriate.
Pump (modified)
Flow
Modified VC-pumps except VC‑P10
4 µL
Modified VC‑P10
6 µL
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...