Modifying the Pump for Specific Applications
Page 248
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
7. Connect the inline filter outlet to the autosampler injection valve
Connecting the Pump and Autosampler (
).
8. In the Chromeleon software, set the
StaticMixer
property to the
value indicated on the inline filter. If the property is not set to the
correct value, the leak tests may not provide reliable results.
9.
With Chromeleon 7.2.8 and later:
In the Chromeleon software, set
the
CapillaryMixer
property to the value indicated on the capillary
mixer.
10. Apply the flow rate of your application and have the pump deliver
for a short time.
11. Inspect the flow connections on the mixer for indications of leakage
before resuming operation.
12. In the Chromeleon software, consider updating the Predictive
Performance information for the static mixer (
MixerChanged
command).
When operating the pump, observe the following
•
Operating the pump with the inline filter reduces the gradient delay
volume but increases the ripple.
•
The maintenance procedures for the static mixer apply equally to
the inline filter (see
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...