5
•
Installation
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 79
Flushing the Seal Wash System from Hand
To flush the seal wash system from hand, follow these steps:
1. Disconnect the tubing from the seal wash detector.
Figure 27: Seal wash detector
No.
Description
1
Seal wash detector
2
Tubing connection port (port labeled
in
)
2. Insert a syringe into the open end of the tubing.
3. To allow the liquid to travel easily through the seal wash system,
press down the lever of the seal wash pump and hold it in that
position with one hand.
4. Pull out the syringe plunger to draw seal wash liquid into the tubing.
5. Release the lever of the seal wash pump when the liquid reaches the
syringe.
6. Remove the syringe and reconnect the tubing to the connection port
on the seal wash detector.
Mind the port labeling. To ensure correct functioning of the seal
wash detector, the tubing needs to be connected to the port labeled
in
.
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...