7
•
Maintenance and Service
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 135
3. Grasp and pull the seal wash detector assembly upward by the
funnel.
4. Push the new seal wash detector assembly into the pump.
The seal wash detector and funnel come as an assembled unit. If you
have to assemble both parts for any reason, push the detector into
the funnel first. Mind the orientation of the funnel.
Figure 35: Installing the seal wash detector assembly
No.
Description
1a
Funnel (bottom side)
1b
Funnel (top side)
2
Seal wash detector
5. Reconnect the tubing to the seal wash detector (port labeled
in
).
6. Connect the detector cable to the
DROP DET
port.
7. Flush the seal wash system, using fresh seal wash liquid (see
Flushing the Seal Wash System (
).