9.2.6.2
Monitoring the Solvent Consumption
TIP
When monitoring the liquid level in the solvent reservoirs (solvent
consumption) as described in this section, the Chromeleon software
calculates
the liquid level. To monitor the liquid level in real time, based
on real
physical
measurement, install the Vanquish Solvent Monitor. The
solvent monitor is available as an option (see
).
For monitoring the solvent consumption, note the following:
•
The two pumps can share the same set of solvent reservoirs or they
can be connected to different sets of reservoirs.
•
Depending on the setup, you can monitor the solvent consumption
for the dual pump as a whole or separately for each pump.
Monitoring the Solvent Consumption for the Dual Pump as a Whole
TIP
It is possible to share solvent reservoirs even if the two pumps are
assigned to different instruments.
1. In the Instrument Configuration Manager, open the dialog box for
the dual pump.
2. On the
Bottles
tab page, select the
Share Eluent Bottles
check box.
3. In the Chromeleon user interface, enter the information that is
required for monitoring (see
Important Operating Parameters
). Enter the information for both pumps.
Notice that the monitoring properties show the same values for both
pumps, for example, the
RemainTime
property.
Monitoring the Solvent Consumption for Each Pump Separately
1. In the Instrument Configuration Manager, open the dialog box for
the dual pump.
2. On the
Bottles
tab page, clear the
Share Eluent Bottles
check box.
3. In the Chromeleon user interface, enter the information that is
required for monitoring (see
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...