6
•
Operation
Page 102
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Parameter
Description
Seal washing
The seal wash system is enabled and cannot be
disabled. As a standard, the pump performs a seal
wash in the following situations:
•
Once per hour
•
Each time the pump is turned on
If required, you can start an additional seal wash cycle
or stop a running cycle:
•
If the seal wash pump is in
Idle
mode, select
Active
to start a wash cycle.
•
If the seal wash pump is in
Active
mode, select
Idle
to stop a running wash cycle.
You can disable the drop counter functionality by
setting
Rear Seal Wash Monitoring
to
Disabled
. Note
that disabling drop detection will not disable the seal
wash system. However, messages relating to the seal
wash system (for example, when the system has run
out of wash liquid) or piston seal leakage will no
longer alert you in the user interface.
Synchronization of the
injection time with the
pump strokes
Relevant for quaternary pump and dual pump only
Synchronizing the injection time with the pump
strokes enhances the retention time precision. For
details, see
Synchronizing the Injection Time with the
.
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...