5
•
Installation
Page 80
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
5.8
Connecting the Pump and Autosampler
Parts required
Capillary to connect the pump to the autosampler, as required for the
flow path
Follow these steps
1. On the connecting capillary, remove the knurls.
2. Guide the connecting capillary through the guide holes in the
housings of both, the pump and autosampler.
Figure 28: Guide holes in the autosampler and pump (left: view from VF‑/
VH-autosampler (example); right: view from pump)
No.
Description
1
Capillary connecting the pump and autosampler:
1a
Capillary end connected to the autosampler injection valve
1b
Capillary end to be connected to pump outlet
2
Guide holes in autosampler housing and pump housing:
2a
Guide hole in autosampler housing (view from above)
2b
Guide hole in pump housing (view from below)
3. Reattach the knurls to the connection capillary.
4. In the pump, connect the capillary to the pump outlet.
5. In the autosampler, connect the capillary to the injection valve.
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...