4. Assemble the solvent line filter, wearing appropriate clean gloves:
a) Place the frit in the filter holder (bottom part).
b) Make sure that the frit is in a level position.
c) Screw the filter top to the filter bottom.
5. Slide the filter holder onto the solvent line.
6. Retighten the reservoir cap hand-tight. Verify that the retaining
guide remains in the hole in the reservoir cap. If it does not, press
the retaining guide into the hole to ensure that the tubing is kept in
place in the cap.
7. To remove any air bubbles from the solvent lines, purge the pump.