6
•
Operation
Page 112
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
5. Cut the solvent line near the reservoir cap with a tubing cutter.
Make sure that the cut is at a right angle to the length of the line.
6. Onto the free end of each solvent line, slide a fitting and a ferrule.
Mind the correct orientation of the fitting and ferrule (see picture).
7. Tighten the solvent lines to the shut off valve.
8. Check the solvent lines across the entire flow path. Make sure that
the lines are not bent, pinched, or squeezed at any point in the flow
path.
9. To remove any air bubbles from the solvent lines, purge the pump.
Figure 31: Solvent line with shut off valve installed
No.
Description
1
Shut off valve
2
Retaining guide