7
•
Maintenance and Service
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 179
7.11 Testing the Pump for Leakage (General Test)
When
After maintenance on the fluidic system of the pump, for example on
the flow connections, pump head, or check valves
TIP
With the dual pump, test each pump separately.
Parts and additional items required
•
Fitting plug, Viper
•
Solvent suitable for your application
Follow these steps
1. Disconnect the capillary on the pump outlet.
2. Close the pump outlet with the fitting plug.
3. In the Chromeleon software, set the upper pressure limit to the
appropriate value:
Pump
Upper Pressure Limit
VC‑pump
60 MPa
VF‑pump
80 MPa
4. Apply a flow of 50 µL/min and use the solvent for the application.
5. Observe the pressure. When the pressure reading is as indicated in
the table, change the flow rate to a lower value.
Pump
Pressure
VC‑pump
Between 30 MPa and 40 MPa
VF‑pump
Between 60 MPa and 70 MPa
6. Have the pump deliver until the following pressure has built up:
Pump
Pressure
VC‑pump
55 MPa
VF‑pump
75 MPa
7. When the pressure has built up, apply the following flow:
Pump
Flow
VC‑P10, VC‑P20,
VC‑P32, VC‑P40
1 µL/min
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...