Alignments: Plasma Cleaning / External Plasma Cleaning
4-16
User Manual
C O N F I D E N T I A L –
limited rights
Feb 2018
Revision A
Plasma Cleaning / External Plasma Cleaning
When repeating the
Tools
menu /
Sample Cleaning
procedure is not effective, click on the
Start / Stop Chamber
Cleaning
button to start / stop the procedure. Its
Duration
is much longer.
If a longer plasma cleaning time is needed, we recommend using several shorter plasma cleaning cycles separated
by pumping the chamber down to HiVac mode. Residuals created during plasma cleaning are pumped away, and the
following plasma cleaning is more effective. Adjust the
Cycle duration
,
Clean cycles
and
Pumping time between cycles
.
Note
If there is an external Evactron plasma cleaner, follow the External Plasma Cleaning alignment instructions and the Evactron
user manual!
W A R N I N G !
Do not use Chamber cleaning when the EDS / WDS / EBSD system is mounted! If you need to clean the system with an
EDS detector mounted, use a 1 mm collimator and do not exceed 30 minutes once a week at most.
Do not exceed 30 minutes of chamber cleaning once a week at most when the DBS detector is installed.
Do not leave sensitive samples (including Au-C resolution test samples) in the chamber during the Chamber Cleaning
(they may be damaged).