Operating Procedures: Microscope Control
5-8
User Manual
C O N F I D E N T I A L –
limited rights
Feb 2018
Revision A
FIGURE 5-5:
X-Ray Cone mounting Procedure
To mount the
X-Ray Cone
follow this procedure:
5.
Push the cone attached in the tool to the final lens pole in its axis direction.
6.
Release the arresting screw, slide the working part down, and pull the tool to the side.
7.
Turn the tool by 180° and push the cone again to ensure it is sealed well.
To dismount the cone, follow steps 5. and 6. in reverse.
Note
Do not touch the cone or the mounting tool by hand. Store it attached to the mounting tool in a clean plastic bag.
8.
Select an appropriate cone and click the
Vacuum
module /
Pump
button. While pumping, choose the highest
specimen point and bring it to the 10 mm Working Distance (yellow line in optical display).
Imaging Onscreen
Continue the procedure:
9.
When the vacuum status is PUMPED (see the status bar), click
on the
Column
module /
Beam On
button to ramp up the
electron beam acceleration voltage.
10.
Select an appropriate column Use case, detector and release
the active display (if paused), where the image appears.
11.
Focus the imaging and
Link Z to FWD
.
12.
Adjust to a suitable magnification. Optimize the imaging using the
Contrast & Brightness, Focusing, Astigmatism
Correction
etc.
5.
X-Ray cone installed
6.
7.