System Options: Uninterruptible Power Supply (UPS)
7-3
C O N F I D E N T I A L
– limited rights
User Manual
Revision A
Feb 2018
Uninterruptible Power Supply (UPS)
If power failures occasionally occur, using the microscope UPS is recommended. It maintains the electron source
section.
In case the microscope system is powered by the UPS and a mains power failure happens, the system starts a 10
minute countdown to switch to the safe mode. The user is informed of the countdown progress in the application
status window. If the mains power is recovered within this time, the countdown is cancelled and nothing happens.
After 10 minutes of continuous power off, a shutdown to the safe mode with the following actions is activated:
•
FEG emission is gently switched off
•
Chamber is vented
•
UI is stopped
•
xT Microscope server is stopped
•
Microscope console is switched off to Standby mode
•
Microscope controller and the support computer (if present) are switched off
W A R N I N G !
Because the Emitter IGP’s are supported by the UPS, some parts of the microscope are still under power.
To return the system to operation, follow the startup procedure (see Chapter 2). When the xT microscope Server is
launched the first time from the safe mode, a dialog is shown to inform a user.
Note
A Supervisor must restart the system after a longer power failure. If the Startup procedure fails, contact an Thermo
Scientific Service Engineer.