System Options: CryoCleanerEC
7-32
User Manual
C O N F I D E N T I A L –
limited rights
Feb 2018
Revision A
CryoCleaner Operation
Once mounted, the Nitrogen vessel can be placed in the Vacuum vessel. Secure the two components by fixing the
clips to the top of the Nitrogen vessel and locking the clips down. Take care that the 'O' ring seal on the Vacuum
vessel is secure when joining the two components together.
Dewar Vessel Refilling
W A R N I N G !
The handling of LN2 should be performed wearing face and hand protection in the form of a face visor and a pair of
thermal protective gloves.
Users must not touch the cold surfaces of the Dewar, as this can result in burns. Use the Safety pliers provided when handling
the Nitrogen Vessel.
1.
Pump the specimen chamber. The Vacuum vessel is pumped along with it.
2.
When the specimen chamber vacuum is ready (Pumped status), partially fill the Dewar, using the funnel (the
plastic cap upside down). Wait until boiling stops.
3.
Then fill the Dewar and place the plastic cap at the top of the CryoCleaner. The volume of liquid Nitrogen
needed is approximately 500 ml.
Note
The LN2 stops boiling very quickly so that no vibration is seen from this device. If the CryoCleaner needs to be used for longer
periods, it can be refilled with LN2.
Before re-filling it, vessel baking is recommended.
FIGURE 7-9:
Plastic Cap Filling Position / Plastic Cap Operation Position
Removing Nitrogen Vessel
W A R N I N G !
Use the Safety pliers provided when handling the Nitrogen vessel.
Removing the Nitrogen vessel depends on the level of contamination found in the specimen chamber. If the level is
unusually high, then the CryoCleaner can work continuously till improvement is seen. Otherwise, normally after
approximately 2 to 3 hours, the Nitrogen vessel can be removed.
Note
We recommend not leaving it inside the vacuum vessel after all the nitrogen has evaporated, because contamination evaporates
back into the chamber.
If the chamber is vented with the Nitrogen vessel filled with LN2 (or still significantly below the ambient temperature), it
should be removed and baked before it is re-used.