System control: Interface elements
2-3
C O N F I D E N T I A L
– limited rights
User Manual
Revision A
Feb 2018
External connectors panel
The External connectors panel is located on the back of
the microscope console.
It is used to connect 3rd party equipment to the
following connectors:
CONTROL SIGNALS / E-BEAM SCAN INPUT / VIDEO
OUT
These connectors are used for the connection of EDX,
WDX and lithography systems, provided by 3rd party
suppliers. Do not connect anything else to these
connectors!
SPECIMEN CURRENT
This connector is used together with the Keithley
picoamper meter (option) equipment.
Ground point
The point is intended for a defined ground connection
between a microscope and a 3rd party module. This
connection is not a safe ground!
ELPHY SCAN & BLANKING IN
Not used.
Mains switchboard
Power off
The system has protection against power failures in the sense that the different components of the system are not
likely to be damaged. However, a power failure is never good for the system. It might affect the ultrahigh vacuum
levels of the columns and the overall stability of the system.
Take sufficient measures to avoid power failures. If a power failure occurs while the instrument is completely
operational, the microscope enters a safe state and the following happens:
•
Electron accelerating voltage is switched off.
•
Electron emission is switched off.
•
The specimen chamber automatically vents gently.
The vacuum in the instrument is no longer supported by running pumps. The column isolating valves close to
save the vacuum in the source area.
•
The microscope computer and the support computer are switched off. The momentary adjustments of all
system parameters (accelerating voltage, magnification, stage positions, etc.) are lost if they were not saved.
Note
If mains failures occur occasionally, using the Uninterruptible power supply (UPS – option) is recommended.