NSG 3040 EMC test system
Decreasing time Td:
ms:
1 ... 5000
s:
1 ... 5
cycle: 1 ... 250 cycles for 50 Hz
1 ... 300 cycles for 60 Hz
Abrupt
Time at reduced voltage Ts:
ms:
1 ... 10000
s:
1 ... 10
cycle: 1 ... 250 cycles for 50 Hz
1 ... 300 cycles for 60 Hz
Increasing time Ti:
ms:
1 ... 5000
s:
1 ... 5
cycle: 1 ... 250 cycles for 50 Hz
1 ... 300 cycles for 60 Hz
Test duration:
s:
1 ... 99’999
min: 1 ... 70’000
pulse: 1 ... 99’999
Continuous
Automatic accessories for power quality test
All automated standard accessories for PQT test provide a convenient means
of reducing the incoming supply voltage.
Once detected, the functions are available in the user interface software. It’s
fully automatic controlled, driven from NSG 3040.
With the automatic step transformer INA 6502 the Uvar settings 0% – 40% –
70% – 80% will appear.
Connecting the single variac VAR 3005-S16 (or VAR 6501) the settings of Uvar
will be possible in volts or % of Uin. Therefore Uin needs to be set first in the
“General” settings menu. Uin in this case is the actual input voltage of the
single variac.
When using the double variac VAR 3005-D16 (or VAR 6502) it is important that
Uin in the “General” setting gets set first before entering the variation screen.
The value of Uin is variable with the double variac.
Содержание NSG 3040
Страница 1: ...1 NSG 3040 EMC TEST SYSTEM USER MANUAL 601 279F ...
Страница 2: ...NSG 3040 EMC test system NSG 3040 EMC TEST SYSTEM USER MANUAL ...
Страница 80: ...80 NSG 3040 EMC test system Explanation of Burst frequency Burst duration and Repetition time parameters ...
Страница 102: ...102 NSG 3040 EMC test system 13 DECLARATION OF CONFORMITY CE ...
Страница 155: ...155 NOTES ...