59
8.3.3 Burst parameters window
Burst parameter setting window
8.3.4 Voltage
Touch the “Polarity” button (ALT in the example) to select test polarity. Polarity
values are:
positive (POS), negative (NEG),
or
alternating (ALT).
On odd pulse number there will be one pulse less in negative then in positive.
Positive pulse will be first executed.
Touch the “Voltage” button (200 V in the example) to enter the test voltage. A
red frame is displayed around the field. The voltage value may be entered using
the wheel or the keypad.
8.3.5 Frequency
Touch the “Frequency” button (5 in the example) to set the test frequency. A
red frame is displayed around the field. The frequency value may be entered
using the wheel or the keypad.
Touch the units button (KHz in the example) to set the frequency unit. Frequency
values are Hz and KHz.
8.3.6 Phase
Touch the Synch/Asynch button (Asynch in the example) to activate the syn-
chronization of test pulses to the EUT mains frequency.