109
Installation - connection to NSG 3040
1) Disconnect instrument power from the mains
2) Remove the 25-way Sub D plug on the rear of the NSG 3040
3) Connect the 25-way Sub D cable to the X2 plug of the VAR 3005
4) Connect the master controller 25-way output to the VAR 3005 X1 plug,
using the system interface cable provided with the VAR 3005
5) Connect the VAR 3005-EUT power out to the NSG 3040 – EUT power input
6) Connect the VAR 3005-EUT power in to mains using the EUT power in
cable delivered with NSG 3040
MAINS should be switched off during installation and
interconnection.
Because of the capacitors in the internal coupler of NSG
3040, earth leakage currents of up to 4 A can occur in
the EUT power supply network. The test system must
therefore be correctly earthed and be powered from a
supply that is not protected by a residual current detec-
tor (RCD).
1) Switch on the VAR 3005 first
2) Switch on the NSG 3040
3) Switch on the EUT power (red switch) if power for the EUT is required.
HW Detection
The VAR 3005 is automatically detected by the NSG 3040 during the booting
process. Its presence is visualised in the system settings screen.