Thermal monitoring of internal backfilter chokes; in case the EUT current
goes up, the integrated fans, which are still in standby mode and at low
EUT currents, will speed up to improve cooling.
In case of intentional or unintentional overloading, the CDN 3043 and 3063
series will automatically switch off EUT power, in order to protect itself
(risk of fire)
Slow Damped Oscillatory Wave immunity testing of 3-Phase equipment needs
a dedicated 3-Phase CDN. Please check your Teseq equipment supplier for
availability.
14.2.1 CDN 3043 - 32 A series
Technical specifications:
NSG 3040 on top of CDN 3043
Voltage ratings:
280 VAC - phase to neutral or phase to ground
480 VAC - phase(s) to phase(s)
up to 125 VDC - full current range
up to 225 VDC - for max. 7 A