118
NSG 3040 EMC test system
Parts description
Plug X3: Remote TEST and EUT power off: with shorting plug. Allows to connect
external door switch or equivalent. Interrupts the 24 V supply of the circuit
breaker contactor which switches EUT supply on/off.
The information will be transmitted to NSG 3040 which will stop the test.
Plug X2: NSG 3040 system interface OUT – to be terminated by interlock plug
or to be linked to another accessory, to X1 plug.
Input voltage:
0 to 250 VAC (not suited for DC
voltages)
Output voltage:
4 steps: 0, 40, 70, 80%
Accuracy:
± 5%
Voltage change with load:
100% output , 0 to 16 A
less than 5%
80% output, 0 to 20 A
less than 5%
70% output, 0 to 23 A
less than 5%
40% output, 0 to 40 A
less than 5%
Output current capability at
230 V input voltage:
16 Arms at 100% output
20 Arms at 80% output
23 Arms at 70% output
40 Arms at 40% output
Voltage selection:
Software driven
EUT power on/off function:
Front panel switch with on indicator
software driven from NSG 3040
Fuse:
16 A, slow blow
Connectors:
Harting type HAN3A
Power Supply:
Selectable 100-110 V, 220-240 V, 15 VA
Size:
460 x 200 x 160 mm
Weight:
15 kg approx.
Output cable length:
2 meter
Input cable:
NSG 3000 standard cable to be used
Control cable:
2 meter - 25 way sub D - twisted pair -
shielded (included in delivery)
Technical specifications INA 6502
Содержание NSG 3040
Страница 1: ...1 NSG 3040 EMC TEST SYSTEM USER MANUAL 601 279F ...
Страница 2: ...NSG 3040 EMC test system NSG 3040 EMC TEST SYSTEM USER MANUAL ...
Страница 80: ...80 NSG 3040 EMC test system Explanation of Burst frequency Burst duration and Repetition time parameters ...
Страница 102: ...102 NSG 3040 EMC test system 13 DECLARATION OF CONFORMITY CE ...
Страница 155: ...155 NOTES ...