5.1 Installation of the NSG 3040 system
The mains power voltage indicated on the instrument must correnspond with
the local supply voltage (mains voltage: 85–265 VAC, universal power unit, mains
frequency: 50–60 Hz).
Mains switch
Fuse-holder with fuses 2 x 3.15 AT
Mains power input
To replace a fuse:
1) Disconnect the mains cable
2) Pull the fuse holder out of the connector
3) Remove the damaged fuse(s)
4) Insert 1 or 2 x 3.15 AT fuses
5) Replace the fuse holder
6) Plug the mains cable into a power outlet with a solid ground connection
7) Switch the system on and operate as instructed in this manual
Mains switch, fuse holder and power input
NOTE: Place the test system so that there is sufficient
free space around the cooling air inlets on both sides
and behind the fan outlet on the rear panel.
NOTE: In case the test system is being used without an
Emergency Stop device, make sure to place the system
such as the operator has quick access to its power
supply switch.