8.6 Power Frequency Magnetic Field testing (-4-8) parameter setting
For proper operation of the plug and play detection
mechanisms it is strongly recommended to power on
first the accessory and then the NSG 3040 main frame.
Powering on the NSG 3040 main frame before the acces-
sories may result in non-detection of accessories.
Power Frequency Magnetic Field test window
Parameter Value
Field strength:
1 to 100 A/m (in 1 A/m steps)
Frequency:
50 Hz/60 Hz
Coil factor:
0.01 to 99.99
Test duration:
s:
1 ... 9’999
min:
1 ... 166
Continuous
More information about variable voltage sources is available in section 14
“Accessories”.