124
NSG 3040 EMC test system
The INA 703 can also be used with the MFO 6501 or 6502 current sources and
the NSG 3000 series of generators to generate supply frequency fields (50 Hz
and 60 Hz) over 120 A/m continuously.
With its multi-turn concept and professional mechanical design features, such
as the U-shaped caster base for convenient positioning at the test table, the
INA 703 is the ideal accessory for magnetic field testing.
In order to meet the pulse waveform required by IEC 61000-4-9, the waveshape
adapter INA 752 needs to be used with NSG 3040 and the INA 701 and INA 702
loop antennas.
The tests are carried out using the so-called immersion method, i.e. the item
under test is placed in the center of the loop. The test is performed in accor-
dance with the IEC 61000-4-8 (mains frequency) or IEC 61000-4-9 (pulsed)
standards for magnetic fields.