27
6.1.3 Surge or SOW output
NSG 3040 models including CWM 3450 (Combination Wave/Surge Module)
These sockets (high, low) connect the surge output signal to an external CDN
or to another external coupling unit.
These sockets are also used to connect the internal generator to the optional
magnetic field coil for tests with pulsed magnetic fields.
The surge output is potential free (floating). The inner conductor of each con-
nector is the surge high and surge low connection respectively, while the outer
conductor (screen) is connected to the NSG 3040’s ground terminal.
NSG 3040-SOW
For the NSG 3040-SOW models, both of these sockets are used to connect the
slow damped oscillatory wave signal to an external coupling device.
The surge output is potential free (floating). The inner conductor of each con-
nector is the surge high and surge low connection respectively, while the outer
conductor (screen) is connected to the NSG 3040’s ground terminal.
6.1.4 Burst output
This socket connects the instrument to an external burst coupling clamp for
capacitive coupled burst tests on data lines, or to an external coupling network
Note:
For NSG 3040-SOW models the Burst output is inactive.
6.1.5 Indicator LEDs
The LEDs switch on and off during the boot period and when errors occur.
LED indicator
Function
Power on:
Instrument/system in operation
Pulse:
Shows the occurrence of pulses or a test event
High voltage active: Shows that high voltage is present in the instrument
EUT-Power on:
Indicates when the EUT power supply is present at
the EUT connector on the front panel
Error:
Indicates that a system error has occurred
Содержание NSG 3040
Страница 1: ...1 NSG 3040 EMC TEST SYSTEM USER MANUAL 601 279F ...
Страница 2: ...NSG 3040 EMC test system NSG 3040 EMC TEST SYSTEM USER MANUAL ...
Страница 80: ...80 NSG 3040 EMC test system Explanation of Burst frequency Burst duration and Repetition time parameters ...
Страница 102: ...102 NSG 3040 EMC test system 13 DECLARATION OF CONFORMITY CE ...
Страница 155: ...155 NOTES ...