143
MD
200
MD
200A
Technical specifications MD 200
Attenuation ratio:
2 ranges: 1:100 and 1:1000
Bandwith:
DC to 10 MHz
Accuracy:
+/2%
Max. input voltage different mode: 7000 V peak
Max. input voltage common mode: 3500 V peak
Input impedance:
10 MΩ/7 pF each side ground
CMRR (typical):
-80 dB at 50 Hz; -60 dB at 20 kHz
Operating temperature:
-10° to + 40C° (14° to 104° F)
Dimensions (LxWxH):
207 x 83 x 38 mm (8.1 x 0.32 x 0.15”)
Connector to scope:
BNC and auxiliary earth lead
Input connectors:
HV alligator clip
Weight:
500 g approx. (1.1 lbs)
Technical specifications MD 200A
Attenuation ratio:
2 ranges: 1:100 and 1:1000
Bandwith:
DC to 10 MHz
Accuracy:
+/2%
Max. input voltage different mode: 7000 V peak
Max. input voltage common mode: 7000 V peak
Input impedance:
10 MΩ/7 pF each side ground
Содержание NSG 3040
Страница 1: ...1 NSG 3040 EMC TEST SYSTEM USER MANUAL 601 279F ...
Страница 2: ...NSG 3040 EMC test system NSG 3040 EMC TEST SYSTEM USER MANUAL ...
Страница 80: ...80 NSG 3040 EMC test system Explanation of Burst frequency Burst duration and Repetition time parameters ...
Страница 102: ...102 NSG 3040 EMC test system 13 DECLARATION OF CONFORMITY CE ...
Страница 155: ...155 NOTES ...