Operation
The parameter field voltage Uvar gets active once an INA 6502 has been
detected by the software. The entry of Uin is by default 230 V and may be set
to other voltages, this can be done in the system/general windows. Uvar can be
selected for one of the 4 available variable voltage levels: 0% - 40% - 70% - 80%
Circuit diagram INA 6502
2.1 Circuit diagram INA 6502
L
EUT
Power
IN
N.C
PE
N
EUT Power ON/OFF
16 AT
Circuit Breaker Option (CIB)
Instr.
Supply
Power
Supply
0% 40% 70% 80%
EUT Power
ON/OFF
Interface Controller
Plug XI - System cable IN
25 Way Sub D – Male
Plug X2 - System cable OUT
25 Way Sub D – Female
X
3
Uin
Uvar
PE
N
EUT
Power
OUT
Remote
Test + EUT
Power OFF
Connection to NSG
MaterController via
System Interface cable
Termination with
Interlock plug
For proper operation of the plug and play detection
mechanisms it is strongly recommended to power on
first the INA 6502 accessory and then the NSG 3040.
Powering on the NSG 3040 main frame before the acces-
sories may result in non-detection of accessories.
The unit has been designed for use in rugged industrial environments. Profes-
sional quality connectors ensure user safety, additional system protection is
provided by a 16 A fuse located in the top panel.
Thanks to the provision of an 80% voltage position and to the large over current
capabilities the step transformer is fully compliant with the latest requirements
called for in IEC 61000-4-11:2004 standard.
Содержание NSG 3040
Страница 1: ...1 NSG 3040 EMC TEST SYSTEM USER MANUAL 601 279F ...
Страница 2: ...NSG 3040 EMC test system NSG 3040 EMC TEST SYSTEM USER MANUAL ...
Страница 80: ...80 NSG 3040 EMC test system Explanation of Burst frequency Burst duration and Repetition time parameters ...
Страница 102: ...102 NSG 3040 EMC test system 13 DECLARATION OF CONFORMITY CE ...
Страница 155: ...155 NOTES ...