24
NSG 3040 EMC test system
5.2 Connecting the system to the ground reference plane
For burst tests, the generator must be placed on a ground reference plane
which is connected to ground. A good high frequency ground connection
between the test system and the ground reference plane (GRP) is absolutely
essential for performing burst tests correctly.
Connect the ground terminal on the front panel of the NSG 3040 to the ground
reference plane using the link and bolts supplied. If a CDN is connected please
refer to section “High frequency ground terminal”.
5.3 Mounting in a 19” rack
When the NSG 3040 test system is combined with other equipment, it can be
useful to mount the instrument in a 19“ rack. The unit is 19” wide and 5U in
height. An optional rack mount INA 166 kit is available.
5.4 Rear ground brackets (INA 165)
Rear ground brackets are optionally available to position the NSG 3040 securely
without damaging the connectors when it must be placed with the rear panel
on the floor with easy access to the touch screen. These brackets guarantee a
solid ground connection to the GRP. The stable housing construction allows the
operator to make use of both back brackets as well as the handles.
NOTE: Your NSG 3040 generator has been delivered with
a correctly rated power supply cable. If the cable needs
to be replaced, the user needs to make sure the new
cable is suited for the rated supply voltage and current.
Содержание NSG 3040
Страница 1: ...1 NSG 3040 EMC TEST SYSTEM USER MANUAL 601 279F ...
Страница 2: ...NSG 3040 EMC test system NSG 3040 EMC TEST SYSTEM USER MANUAL ...
Страница 80: ...80 NSG 3040 EMC test system Explanation of Burst frequency Burst duration and Repetition time parameters ...
Страница 102: ...102 NSG 3040 EMC test system 13 DECLARATION OF CONFORMITY CE ...
Страница 155: ...155 NOTES ...