38
NSG 3040 EMC test system
Expert mode
Change the “Expert mode” status to “Active” to change parameters during a
running test. When the button is set to “Off” parameters can be changed only
when the NSG 3040 is in Stop mode. Expert mode is only available for burst
pulses (EFT).
Voltage U in
This button is active only when an optional automated VAR 6502 or VAR 3005
variac is connected to the NSG 3040. The value entered in this field is the
voltage measured at the mains socket and is used as the 100% reference point
for voltage variation tests.
Touch the “Voltage Uin” button (“230” in the example). Use the wheel or keypad
to set the input voltage.
Uin setting will be saved and is valid for all following tests. Uin are changeable
via WIN 3000 (dialogue) to be used in sequence mode.
EXIT
Touch the “Exit” button to return to the system window without saving settings.
EUT ON/OFF
The “EUT ON/OFF” button is used only when an option with a built-in EUT switch,
such as an INA 6502, a CDN 3043 or a VAR 3005, is connected to the NSG 3040.
The NSG 3040 itself does not have an EUT switch. Touching the button will turn
the EUT switch on or off.
FACTORY SETTINGS
Touch the “FACTORY SETTINGS” button to reset the properties associated with
each of the buttons in the general settings window to the original factory set-
tings.
OK
Touch the “OK” button to save all settings and return to the system window.
Содержание NSG 3040
Страница 1: ...1 NSG 3040 EMC TEST SYSTEM USER MANUAL 601 279F ...
Страница 2: ...NSG 3040 EMC test system NSG 3040 EMC TEST SYSTEM USER MANUAL ...
Страница 80: ...80 NSG 3040 EMC test system Explanation of Burst frequency Burst duration and Repetition time parameters ...
Страница 102: ...102 NSG 3040 EMC test system 13 DECLARATION OF CONFORMITY CE ...
Страница 155: ...155 NOTES ...