81
8.8.2 Coupling
Touch the “Coupling mode” button in the top row (“Surge output” in the example
above) to select
Surge Output
,
Manual CDN
or
EUT Power Output
.
Surge Output
Surge Output must be selected if an external capacitive coupling clamp (e.g.
CDN 3425) is connected to the NSG 3040-SOW.
Manual CDN
Manual CDN must be selected if an external manual coupling/decoupling device
is connected to the NSG 3040-SOW. The factory setting for Manual CDN is the
same as for Surge Output.
EUT Power Output
With EUT Power Output setting the Slow Damped Oscillatory pulses are directly
applied on the EUT power output plug on the front panel of the NSG 3040-SOW.
Touch the coupling line selection field (- - - in the example) to display the cou-
pling selection window.
Touch the individual “High output coupling line” buttons (L and N in the example)
to select an open or closed relay.
Touch the individual “Low output coupling line” buttons (PE in the example) to
select an open or closed relay.
Touch “OK” to enable the coupling selection and close the window.
Touch “Cancel” to close the window without saving the coupling selection.
Touch “Show Graphics” to display a graphical example of the coupling selection.
Coupling selection window