ADD STEP
Multi-step tests can be programmed manually in the test parameters window
using the “Add Step” button.
Touch the “Add Step” button create a new step with the values currently dis-
played in the Test parameters window. The user can program a maximum of
10 test steps.
When the first test step is programmed, “Test Step 1/X” is displayed in the upper
right corner, and the step can no longer be changed from the Test parameters
window.
To change a step, the user must first delete it using the “Show Step” button,
then use “Add Step” to re-enter the step.
Refer to sections 8.3 - 8.9 for detailed information on setting parameters for
specific types of tests.
EXPERT MODE
The “Expert Mode” button can be used only if “Expert Mode” is set to “On” in
the System/General settings window (see section 7.3). The unique expert Mode
is a fast and effective method to determine critical threshold values of a device
under test (EUT).
Touch the “Expert Mode” button to manually adjust test parameters using the
wheel while a test is in progress.
A selected test can be started. During run mode the changeable parameter can
be touched, the value window is highlighted with a red frame, like the voltage
frame shown in the examples above. The value can now be changed via wheel
and by pressing again the “START” button, the value will be accepted and on
the pulse output the new value is displayed.
Содержание NSG 3040
Страница 1: ...1 NSG 3040 EMC TEST SYSTEM USER MANUAL 601 279F ...
Страница 2: ...NSG 3040 EMC test system NSG 3040 EMC TEST SYSTEM USER MANUAL ...
Страница 80: ...80 NSG 3040 EMC test system Explanation of Burst frequency Burst duration and Repetition time parameters ...
Страница 102: ...102 NSG 3040 EMC test system 13 DECLARATION OF CONFORMITY CE ...
Страница 155: ...155 NOTES ...