126
NSG 3040 EMC test system
Two safety banana sockets provide a convenient means to connect the loop
antenna, two other ones (shorted by a jumper) to connect an external amp-
meter for monitoring the generated current, as the fi eld generated in the loop
antenna is directly proportional to the current fl owing through it:
Field strength (A/m)H = Cf x I
Where H is the generated fi eld, Cf the coil factor, I the current fl owing through
the loop.
The unit has been designed for use in rugged industrial environments. Pro-
fessional quality connectors ensure user safety, additional system protection
is provided by a temperature sensor located on the heatsink of the power
amplifi er.
MFO 6501 is designed to drive INDUCTIVE LOADS ONLY,
as magnetic fi eld loops.
Connecting capacitive loads will destroy the Instrument
Circuit diagram MFO 6501
Содержание NSG 3040