62
NSG 3040 EMC test system
Parameter Value
Pulse amplitude:
± 200 V to 4.8 kV (in 1 V steps) - open circuit
± 100 V to 2.4 kV (50 Ω matching system)
Voltage step:
1 V/10 V/100 V
Polarity:
Positive/negative/alternate
Frequency:
Hz:
100 ... 99’999
kHz:
1 ... 1’000
Phase:
Asynchronous, synchronous 0° to 359º (in 1º steps)
Coupling:
ANSI/IEC/external/manual
Burst time:
μs:
1 ... 99’999
ms:
1 ... 99’999
s:
1 ... 1’999
Spike:
1 ... 1000
Repetition time:
ms:
1 ... 99’999
s:
1 ... 4’200 (70 min)
Test duration:
s:
1 ... 99’999
min:
1 ... 99’999
h:
1 ... 1’000
Continuous
8.3.11 Burst generator technical data
8.3.12 Derating
Some parameter combinations will not be accepted due to the power limitation
of the HV power supply. The following error message will be displayed when an
invalid combination of parameters is entered:
Invalid parameter error message