26
NSG 3040 EMC test system
Note: For DC power supply L = positive (+), N = negative (-).
The pins in the connector are designed for a maximum current of 16 A.
N
L
WARNING - Never attempt to connect or disconnect an
EUT while a test is being performed.
EUT output connection
6.1.2 High frequency ground terminal
This terminal provides a solid high frequency ground connection point to the
test system. If an external CDN is connected then the ground strap must be
connected from the CDN to the reference ground plane. There is no need to
connect the ground connector from the generator itself, since the burst con-
nector provides the reference ground from the generator to the CDN.
Reference ground terminal
The NSG 3040 can be efficiently connected to the GRP using the ground strap
supplied with the system.
This ground link must be used for burst tests to obtain reproducible test results.