130
NSG 3040 EMC test system
Two safety banana sockets (red and black) provide a convenient means to
connect the loop antenna, two other ones (green - shorted by jumper) to
connect an external amp-meter for verifying (or to calibrate) the generated
current, as the field generated in the loop antenna is directly proportional to
the current flowing through it:
Field strength (A/m) H = Cf x I
Where H is the generated field, Cf the coil factor, I the current flowing through
the loop.
The unit has been designed for use in rugged industrial environments. Pro-
fessional quality connectors ensure user safety, additional system protection
is provided by a temperature sensor located on the heatsink of the power
amplifier.
MFO 6502 is designed to drive INDUCTIVE LOADS ONLY,
as magnetic field loops.
The coil factor of the used loop antenna is to be entered in the respective
field, then the user will setup his tests directly in A/m, the software makes the
calculation and drives the MFO 6502 to generate the right current through the
loop antenna.
Connecting capacitive loads will destroy the Instrument