127
Technical specifications MFO 6501
Parameter
Value
Total harmonic distortion (THD)*:
< 8% (nominal <3,5% at full range)
Frequency:
Selectable 50 and 60 Hz +/- 3%
Range low**:
80 to 400 mA into INA 702 (Cf = 9.8)
=> allows 0.8 to 4 A/m
80 to 440 mA into INA 701 (Cf =
0.89) => allows 0.08 to 0.4 A/m
INTO INA 703 ….
Range high**:
200 mA to 4.1 A into INA 702 (Cf = 9.8)
=> allows 2 - 40 A/m
200 mA to 4.1 A Into INA 701 (Cf = 0.89)
=> allows 0.2 – 3.6 A/m
Supply voltage:
90 to 240 V
Power consumption:
< 150 W
Operating temperature:
5° – 40°C
Overload protection:
By temperature sensor on power stage
Weight:
4 kg approx.
Dimensions:
195 x 180 x 380 mm
* Typical, for the full range from standard level 1 (lowest standard
level) to full range (level X)
** Current adjustment through customer provided amp-meter.
Installation
The equipment should be switched off during installa-
tion and interconnection.
1) Connect MFO 6501 to INA 701, 702 or 703 loop
2) In case of use of INA 702 insure that “power“ plug is fitted
3) Connect MFO 6501 to mains
4) Connect external amp-meter
5) Switch on mains power
6) Adjust the required current through the loop using the knob