114
NSG 3040 EMC test system
Input voltage:
0 to 250 VAC (not suited for DC
voltages)
Output voltage:
4 steps: 0, 40, 70, 80%
Accuracy:
± 5%
Voltage change with load:
100% output , 0 to 16 A
less than 5%
80% output, 0 to 20 A
less than 5%
70% output, 0 to 23 A
less than 5 %
40% output, 0 to 40 A
less than 5 %
Output current capability at
230 V input voltage:
16 Arms at 100% output
20 Arms at 80% output
23 Arms at 70% output
40 Arms at 40% output
Voltage selection:
Friont panel rotary switch
EUT power on/off function:
Front panel switch with on indicator
Fuse:
16 A, slow blow
Connectors:
Harting type HAN3A
Power Supply:
Selectable 100-110 V, 220-240 V, 15 VA
Size:
150 x 180 x 360 mm
Weight:
12 kg approx.
Output cable length:
2 meter
Input cable:
NSG 3000 standard cable to be used
Technical specifications INA 6501
Installation - connection to NSG 3000 series
The equipment should be switched off during installation and inter-
connection.
1) Connect INA 6501 - EUT power out to NSG 3040 - EUT power input
2) Connect INA 6501 - EUT power in to mains using EUT power in cable
(delivered with NSG 3040)
Содержание NSG 3040
Страница 1: ...1 NSG 3040 EMC TEST SYSTEM USER MANUAL 601 279F ...
Страница 2: ...NSG 3040 EMC test system NSG 3040 EMC TEST SYSTEM USER MANUAL ...
Страница 80: ...80 NSG 3040 EMC test system Explanation of Burst frequency Burst duration and Repetition time parameters ...
Страница 102: ...102 NSG 3040 EMC test system 13 DECLARATION OF CONFORMITY CE ...
Страница 155: ...155 NOTES ...