113
The few control elements are readily accessible on the front panel. An EUT
power on/off switch with a power on indicator and a well-proportioned rotary
switch to select the required voltage ensure easy and intuitive operation. The
unit has been designed for use in rugged industrial environments. Professional
quality connectors ensure user safety, additional system protection is provided
by a 16 A fuse located in the front panel.
Thanks to the provision of an 80% voltage position and to the large overcurrent
capabilities the step transformer is fully compliant with the latest requirements
called for in IEC 61000-4-11: 2004 standard.
Care has to be taken in case of use in standing posi-
tion, as the stability is limited. The cabling connecting
INA 6501 to mains and Modula presents a risk of being
unvoluntarly caught by the users which could cause the
INA 6501 to fall down.
It is fitted with carrying handles as part of its overall good ergonomic design,
which makes for ease of handling. Further, the unit may be used in any of three
operating positions; laying or standing on a work bench, or for more permanent
applications, it can be wall-mounted.
1.1 Circuit diagram INA 6501
INA 6501 Step transformer
F1
S1
1 black
2 blue
S1
PE
Ye/Gr
Ye/Gr
blue 2
brown 3
black 3
L
N
PE
L
Lvar.
N
80%
70%
40%
0%
Содержание NSG 3040
Страница 1: ...1 NSG 3040 EMC TEST SYSTEM USER MANUAL 601 279F ...
Страница 2: ...NSG 3040 EMC test system NSG 3040 EMC TEST SYSTEM USER MANUAL ...
Страница 80: ...80 NSG 3040 EMC test system Explanation of Burst frequency Burst duration and Repetition time parameters ...
Страница 102: ...102 NSG 3040 EMC test system 13 DECLARATION OF CONFORMITY CE ...
Страница 155: ...155 NOTES ...