293
May 2016
Reference Manual
00809-0100-4530, Rev DD
Figure B-2. Installation Drawing for Hazardous Location Installation of Intrinsically Safe CSA Approved Apparatus
SME-5514
0644
1
ISSUE
CH. ORDER No
WEEK
ISSUE
CH. ORDER No
WEEK
ISSUE
CH. ORDER No
WEEK
ISSUE
CH. ORDER No
WEEK
A3
ORIGINAL
SIZE
GU-LN
0644
5300
GU-PO
6
P
D
F
9240 030-937
9240 0
30-937
0644
INSTALLATION DRAWING
for hazardous location installation
of CSA approved apparatus
3
1 / 1
ISSUED BY
APPROVED BY
WEEK
WEEK
PRODUCT CODE
DOC. TYPE
FILE
TITLE
DWG NO.
ISSUE
SHEET
SCALE
1:1
1 ST ANGLE
FINISH, UNLESS
OTHERWISE STATED:
ALL DIMENSIONS ARE IN MILLIMETRES.
The copy
right/ownership of this document is and will remain ours.
The docu
ment must not be used without our authorization or brought
to the knowle
dge of a third party. Contravention will be prosecuted.
Rosemount Tank Radar AB, Sweden
A
C
C
OCIA
TED APP
ARA
TUS
BARRIER
POWER
SUPPL
Y
HAZARDOUS L
O
CA
TION
NON-HAZARDOUS L
O
CA
TION
ROSEMOUNT 5300 SERIES
EX-
CERT
IF
IE
D
P
RODUCT
.
N
o
m
odi
fi
cat
ions
per
m
it
te
d
w
it
hout
re
fe
re
nce
to
the
E
x
-cer
ti
fy
in
g
A
ut
hor
it
ies.
Intrins
ic
a
lly
S
af
e
E
x
ia
Clas
s
I,
D
iv
is
io
n
1
,
G
roups
A,B,
C
a
nd
D,
T
e
mperature Code T
4
:
M
o
d
e
l
E
n
tit
y
P
aram
et
er
s
A
m
b
ie
n
t
T
e
mp
erat
u
re
Li
mi
ts
4-20
m
A
/HART
IS
M
od
el
Vm
ax
<
=
30V,
Im
ax
<
=
13
0
m
A
Pi
<
=
1
W
,
C
i
=
7.3 nF
, Li = 0 uH
-5
0<
=T
a<
=7
0d
e
g
C
F
iel
dbus
IS
M
o
d
e
l
V
m
a
x
<
=
3
0V,
Im
a
x
<
=
3
0
0
m
A
Pi
<
=
1.
3
W
,
C
i
=
0,
L
i
=
0
u
H
-5
0<
=T
a<
=6
0d
e
g
C
F
iel
dbus
F
ISCO
IS
M
ode
l
V
m
a
x
<
=
17.5V, Imax <= 380 mA
Pi
<
=
5.
32
W
,
Ci
=
0
,
L
i
=
0
u
H
-5
0<
=T
a<
=6
0d
e
g
C
SME-5983
0840
2
SME-7063
3
1123
ENT
IT
Y
CO
NCEPT
A
P
P
R
O
V
A
L
T
h
e
E
nti
ty
c
o
n
c
ept
al
lo
w
s
in
ter
c
onn
ec
tio
n
of
in
tr
ins
ic
a
lly
s
af
e
a
ppar
atus
to
as
s
o
c
iated
app
a
ra
tus
not
specifically ex
am
ined
in
c
o
m
b
ination
a
s
a
s
y
s
tem
.
T
he
appro
v
e
d
v
a
lues
of
m
a
ximum
open
c
irc
u
it
v
o
ltag
e
(Voc)
and maximum s
h
ort
c
irc
u
it
c
u
rre
n
t
(Isc)
a
n
d
m
aximum output power (or Voc x Isc / 4), for the associated apparatus
must be less than or
eq
ua
l
to
the
m
a
x
im
u
m
s
a
fe
input
vo
lt
age
(U
i),
m
a
x
im
u
m
s
a
fe
inp
ut
c
u
rren
t
(Ii),
an
d
m
ax
im
u
m
safe inp
u
t
p
o
w
er
(P
i)
of
the
intr
in
s
ic
all
y
s
a
fe
ap
parat
us
.
In
a
ddit
io
n,
th
e
a
ppro
v
e
d
m
aximum all
o
w
a
ble connected
c
apac
ita
n
c
e
(Ca)
of
th
e
a
s
s
o
c
iate
d
a
p
par
a
tus
m
u
s
t
be
greater
than
the
s
um
of
the
int
erc
onnec
ting
c
abl
e
capacitance
a
n
d
the
u
nprot
ec
ted
intern
a
l
c
apac
it
anc
e
(Ci)
o
f
the
intr
ins
ic
a
lly
s
af
e
a
pparat
u
s
,
an
d
th
e
approved maximum allowable
connected ind
u
c
tanc
e
(La)
o
f
the
a
s
s
oc
iate
d
a
ppara
tus
m
u
s
t
be
grea
ter
th
a
n the sum of the interconnecting cable
induc
ta
nc
e
a
nd
th
e
u
n
p
rotec
te
d
in
tern
al
in
du
c
tanc
e
(Li)
o
f
the intrinsically safe apparatus.
Notes
:
1. Entity parameters listed (for HART/Fieldbus Model) apply only to associated apparatus with linear output.
2. Control equipment connected to the barrier must not use or generate more than 250 Vrms or Vdc.
3. Connect supply wires to the appropriate terminals as indicated on the terminal block and in the
installation documents.
4. Installations should be in accordance with ANSI/ISA-RP12.6 "Installations of Intrinsically Safe Systems
for Hazardous Locations" and the Canadian Electric Code.
5. Product options bearing the DUAL SEAL marking on the label meets the Dual Seal requirements of ANSI/ISA
12.27.01. No additional process sealing is required. For the in-service limits applicable to a specific model, see
Process Pressure/Temperature range in Appendix A of the Reference manual.
Содержание 5300 Series
Страница 2: ......
Страница 12: ...Reference Manual 00809 0100 4530 Rev DD Contents May 2016 xii Contents ...
Страница 94: ...80 Reference Manual 00809 0100 4530 Rev DD Electrical Installation May 2016 Electrical Installation ...
Страница 162: ...148 Reference Manual 00809 0100 4530 Rev DD Operation May 2016 Operation ...
Страница 214: ...200 Reference Manual 00809 0100 4530 Rev DD Service and Troubleshooting May 2016 Service and Troubleshooting ...
Страница 310: ...Product Certifications May 2016 Reference Manual 00809 0100 4530 Rev DD 296 Product Certifications ...
Страница 358: ...Level Transducer Block May 2016 Reference Manual 00809 0100 4530 Rev DD 344 Level Transducer Block ...
Страница 431: ...5 Reference Manual 00809 0100 4530 Rev DD May 2016 ...